  • 20L capacity
  • Temperature range: +5 ℃ above ambient temperature to +300 ℃
  • APT.line™ preheating chamber technology
  • Natural convection
  • Adjustable exhaust air flap
  • Controller with timer functions
  • 2 chrome-plated racks
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Uniform drying conditions thanks to APT.line™ technology
  • Identical test conditions throughout the chamber interior independent of sample size and quantity
  • Outstanding thermal insulation saves operation costs
Go to Product Description

ASK A QUESTION

Please fill in the form below to ask a question about this product:

  100% Secure Shopping

100% Secure Shopping & Payment Gateway

  Express Delivery

We aim to ship all orders within 1-3 working days.

  Hassle Free Returns

Not 100% happy with your order? find out more about our returns policy.

BINDER ED 23 drying chambers are designed for routine drying and sterilization tasks up to +300 ℃. Thanks to natural convection, all thermal processes which use this drying chamber are highly efficient. The ED series ensures fast, uniform drying. BINDER drying chambers are renowned for their high quality and reliability.

Benefits

  • 20L capacity
  • Temperature range: +5 ℃ above ambient temperature to +300 ℃
  • APT.line™ preheating chamber technology
  • Natural convection
  • Adjustable exhaust air flap
  • Controller with timer functions
  • 2 chrome-plated racks
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Uniform drying conditions thanks to APT.line™ technology
  • Identical test conditions throughout the chamber interior independent of sample size and quantity
  • Outstanding thermal insulation saves operation costs
Technical Specifications
Designation

ED023-230V*

Option model Standard
Order number 9010-0190
Data
Series ED
Performance Data Temperature
Heating-up time to 150 ℃ [min] 28
Recovery time after 30 seconds door open at 150 ℃ [min] 5
Temperature fluctuation at 150 ℃ [± K] 0.5
Temperature range +5 ℃ above ambient temperature to 300 ℃
Temperature variation at 150 ℃ [± K] 2.8
Air change data
Air change (approx.) at 150 ℃ [x/h] 13
Electrical data
Nominal power [kW] 0.8
Phase (Nominal voltage) 1~
Power frequency [Hz] 50/60
Rated Voltage [V] 230
Unit fuse [A] 10
Doors
Unit doors 1
Housing dimensions, not incl. fittings and connections
Depth net [mm] 520
Height net [mm] 495
Width net [mm] 435
Internal Dimensions
Depth [mm] 300
Height [mm] 330
Width [mm] 222
Measures
Interior volume [L] 20
Load per rack [kg] 12
Net weight of the unit (empty) [kg] 27
Permitted load [kg] 25
Wall clearance back [mm] 100
Wall clearance sidewise [mm] 100
Environment-specific data
Energy consumption at 150 ℃ [Wh/h] 148
Fixtures
Number of shelves (std./max.) 2/4
 
*All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 ℃ ±3 ℃ and a power supply voltage fluctuation of ±10%. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.
BIN-9010-0190
50 Items

Our standard delivery is 3-5 working days (where products are available) Monday to Friday, excluding public holidays when you order by 2pm. Details of our returns and cancellation policy can be found in our Returns Policy & Section 11 of our Terms and Conditions. For more information please contact us on 01924 444 577 or Email Us.